EN
Changing pressure | 28
10 Changing pressure
WARNINGS
The settings may only be changed by the physician or on medical orders.
For patients who are to be treated by gravity drainage (= water
seal), the gravity mode can be activated. This mode corresponds
to a pressure of -0.8 kPa / -8 cmH
2
O / -8 mbar / -6 mmHg. To acti-
vate the gravity mode press the [ ] button and confirm with
«OK».
Gravity mode
1.1 Press [
] and [ ]
simultaneously. The
pressure is indicated on
the display.
1.2 Set the desired pressure
with [
] and [ ],
and confirm with «OK».
Press simultaneously
Change pressure Important: Thopaz is running (Data- or Graph mode).