Troubleshooting and alarms | 34
Safety related tip
Canister full
If secretions line the walls of the canister, the warning «canister fulmay be activated
prematurely. The warning «canister ful is triggered only once. After cancelling the
warning with the selection buttons, it is deactivated and a symbol [
] flashes in the
run mode. After deactivation of the warning, the user is responsible for checking the
canister lling level. If the hydrophilic overow/ bacteria lter gets clogged by fluids, the
alarm «System clogge is triggered.
The warning is reactivated under the following conditions:
if Thopaz is switched to standby and then on again for example, on replacing
the canister.
if the canister full sensor is not activated for at least 10 minutes.
Safety related tip
The acoustic alarm «Leak in system - 301» is deactivated for approximately 6 minutes
after switching Thopaz on.