Replace canister | 26
9 Replace canister
WARNINGS
The canister is replaced on the basis of a visual check or according to the instructions on
the display (warning signal). Wear gloves for all operations.
The warning «canister full» is triggered only once. After cancelling the warning with
the selection buttons, it is deactivated and a symbol [
] flashes in the run mode. After
deactivation of the warning, the user is responsible for checking the canister filling level.
2. Close patient tubing with the tube clamp.
4. 4.1
Release the canister by
pressing the release button.
4.2 Remove canister.
5.
6. 6.1 Push the canister into
Thopaz until you hear
a click.
click
3. 3.1 Switch to Standby mode
by pressing «Standby» for
longer than 3 seconds.
1. Prepare sterile canister (with/without solidier).
5.1 Unpack the canister.
5.2 Position the openings at
the top and attach the
bottom part of the canister
to Thopaz.
Safety related tip
Using a 2 L canister
refer to chapter 8.
Canister change