Preparation for use / operating instructions | 22
Carry out functional
check
1. 1.1 Seal off the conical con-
nector of the patient tub-
ing with your thumb
(through the packaging).
The tubing must not be
kinked otherwise the mea-
suring tube will be closed.
2. 2.1 Switch on pressure by
pressing «on» – pressure
is built up.
2.2 Check flow value.
Flow decreases:
Thopaz is ready for use.
Flow does not decrease:
Leak in the system; check tubing connections. Check pres-
sure relief value for tight connection. Replace canister and/
or tubing if necessary.
3. 3.1 Switch to Standby mode
by pressing «Standby» for
longer than 3 seconds.
Thopaz is in Standby mode and is ready for use with
the factory settings.
(To change factory settings please refer to chapter
11-Changing settings)
4.
Connect the Thopaz system to the patients catheter (drain)
according to hospital guideline. To start the therapy press «on».
5.
8.5
Safety related tip
The acoustic alarm function «Leak in system – 301» is deactivated for approximately
6 minutes after Thopaz switching on.