Troubleshooting and alarms | 32
13 Troubleshooting and alarms
beep
Warning
Pressure is maintained.
beep
Alarm
Pressure is interrupted.
beep
Internal error
Problem cannot be
rectied by the user.
The error number is
displayed.
Example:
Problem described on display
Troubleshooting instructions
are displayed
Error number
Press simultaneously, then follow the instructions on
the display
to acknowledge an acoustic alarm
Pump not running
Check if:
the mains voltage is correct and if the mains plug is inserted correctly into the xed
mains socket.
Thopaz is switched on.
Description
Thopaz distinguishes between warnings, alarms and internal errors. If Thopaz detects any of
these situations, an acoustic warning signal sounds and a description of
the problem appears
on the display. By pressing the two selection buttons simultaneously, the acoustic alarm is
suppressed for 60 seconds and the troubleshooting instructions appear on the display.