7 | Warnings and safety instructions
WARNINGS
Consult this manual prior to use.
Data transfer via USB is not allowed during therapy.
Before cleaning the device, pull the plug out of the xed mains socket.
Do not use Thopaz in MRT (Magnetic Resonance Tomography).
Do not dry Thopaz with microwaves.
The device is not suitable for use while bathing, showering or in a hazardous explosive
environment.
The pressure range to be se
t must be determined by a physician in accordance with
the age and weight of the patient.
The secretions gathered must be monitored more regularly or frequently in accordance
with the age and weight of the patient.
To ensure Thopaz can reduce the pressure, it is essential to wait 30 seconds between
taking the sample from the sample port and unclamping the patient catheter (drain).
Do not use Thopaz if the drainage th
erapy indicates a pressure greater than Thopaz
max. pressure range of -10 kPa.
Do not use Thopaz if the drainage therapy indicates a ow rate greater than Thopaz
max. ow capacity of 5 L/min.
Do not use Thopaz if the drainage therapy indicates no pressure should be applied to
the patient.
The catheter/connector interface is a location where clotting may occur.
We recommend regular monitoring of this interface and appr
opriate removal
procedure in case of occurance.
Do not connect bilateral thoracic drains to one Thopaz unit. In such cases, the use of
two Thopaz units is recommended.
The use of Thopaz for any other indication than intended is neither desired nor allowed.
CAUTIONS
No modication of all articles related to the Thopaz system is allowed.
Before connecting Thopaz to the power supply, please verify that the supply v
oltage
corresponds to that given on the device specication plate.
Incorrect use can cause pain and injury to the patient.
Do not use sterile accessories when the sterile packaging is damaged.
For patients for whom a breakdown of the device can lead to a critical situation
a replacement device must always be available.
Caution: Federal US law restricts this device to sale by or on the order of a health
pro
fessional.