Mark V ProVis Injection System
16 - 36
Line Frequency Checkout
Perform this checkout procedure after adjusting the line frequency
according to the procedure on the previous page.
1. Remove power from the injector. Remove the card extender, then
insert the Power Drive Card directly into the chassis, and recon
-
nect the AC wires to the SCR bridge. Connect the injector head,
then apply power to the injector.
2. Reverse the plunger completely. Activate forward load. The
plunger should reach the 0 ml mark +1 ml, within 8-30 seconds.
3. Reverse the plunger completely. Set up an injection of 1 ml/sec for
150 ml. Arm and inject. The injection should be smooth, with no
interruptions or error messages.
4. Reverse the plunger completely. Set up an injection of 50 ml/sec
for 150 ml. Arm and inject. The injection should be smooth, with
no interruptions or error messages.
5. (Pedestal Systems only) Re-assemble the injector following the
procedures outlined in this section, as required.
6. (Rackmount Systems only) Tilt the Front Panel upward, then
secure by tightening the two captive screws.