Troubleshooting Guide
15 - 5
Test Point / Description Typical Values
23 PLNG. MOT. VOLTS LOW Less than 1 VDC while injecting
Plunger Motor VDC Low Side or forward load. +1 VDC with
1200 PSI. Reverse load 7-15
VDC.
24 FLOW SELECTED Measure in ml/sec - 60 ml
Flow Selected syringe, +0.38 VDC/ml. 150 ml
syringe, +0.16 VDC/ml (+6.4
VDC at 40 ml/sec). 200 ml
syringe, +0.12 VDC (+4.8 VDC
at 40 ml/sec). In ml/min or ml/hr,
less than 0.2 VDC.
25 150/200 ENABLE 60 ml = 5 VDC
150/200 Enable 150 ml = 0 VDC
200 ml = 5 VDC
26 F/C START F/C Relay Open = 26 VDC
Film Changer Start F/C Relay Closed = 0 VDC
27 ARM LT. ENABLE Light off = 0 VDC
Arm Light Enable Light on = 5 VDC
Flashing = Pulses 0 to 5 VDC
28 SYSTEM OVERRIDE Standby = 0 VDC
System Override Active state = 5 VDC
29 AUX MONITOR Standby = -14 VDC
Aux. Monitor Active state = 13 VDC
30 +26V +26 VDC