Troubleshooting Guide
15 - 15
ENTER DELAY
The Delta 7 button was pressed. The injector is now ready for an X-
RAY or INJECT delay to be entered.
ENTER DURATION
The Delta 5 button was pressed. The injector is now ready for an
Injection Duration to be entered.
ENTER FLOW RATE
The Delta 2 button was pressed. The injector is now ready for a Flow
Rate value to be entered.
NOTE: Maximum Flow Rate of the 60 ml syringe system is 20.
ENTER PPI NUMBER XX, THEN STORE
The STORE button was pressed; displays the number of the next
unused program storage location. Enter the number shown (or any
number less than 50), then press STORE.
ENTER PRESSURE
The Delta 6 button was pressed. The injector is now ready for a Pres-
sure Limit to be entered.
ENTER PROGRAM NUMBER THEN RECALL
The RECALL button was pressed. Press the number buttons to enter
the program desired, then press the RECALL button again to enter.
ENTER PSI=1,KG=2,KPA=3,ATU=4
Program 51 was recalled to change the Pressure scale. Enter the cor-
responding number for the scale desired.
ENTER RISE/FALL TIME
Delta 1 was pressed. The injector is ready for the user to enter a rise
or fall time.
ENTER SYRINGE SIZE: 60=1, 150=2
Displayed in response to RECALL, 50, in order change the pro-
grammed syringe size; or, displayed after the head selection step in
the RECALL, 65 procedure. Enter the corresponding number of the
syringe to be used in injection procedures.
ENTER SYRINGE SIZE: 150=2, 200=3
Displayed in response to RECALL, 50, in order change the pro-
grammed syringe size; or, displayed after the head selection step in
the RECALL, 65 procedure. Enter the corresponding number of the
syringe to be used in injection procedures.