Mark V ProVis Injection System
15 - 24
** PPI MEMORY COMPARE ERROR
A fault in the PPI memory; the primary memory data does not match
backup data. Remove power from the injector for 10 seconds, then
turn it on and try again.
NOTE: (Pedestal systems only) Pressing the PPI Memory
Reset Switch will erase all programs in PPI memory.
Service Procedure:
1. Clear the PPI memory by pressing the PPI Memory Reset
Switch (S1) on the rear connector panel, then remove and
reapply power to the injector.
2. If the condition occurs repeatedly, replace the PPI card.
PPI MEMORY FILLED
No additional programs can be stored unless an existing program is
cleared first. Delete stored programs which are not, or seldom, used.
PPI NUMBER XX NOT USED
Displayed in response to attempting to recall a Pre-Programmed
Injection that does not exist at the selected location. Scroll through the
programs to search for program desired, or create a new program.
PPI PANEL CHECKSUM ERROR
A fault exists in the PPI memory; the checkless does not match the
parameters displayed on the Control Panel. Press RESET to clear the
fault. If the message appears again, remove and reapply power.
Service Procedure:
1. Clear the PPI memory by pressing the PPI Reset Switch
(S1) on the rear connector panel, then remove and reapply
power to the injector.
2. If the condition occurs repeatedly, replace the PPI card.
3. If the condition persists, replace the CPU Card.
PPI: PRIMARY MEMORY
The primary memory on the PPI Card failed a read/write exercise.
Remove and reapply power, then attempt to operate the unit again.
NOTE:Pressing the PPI Memory Reset Switch will erase all
programs in PPI memory.
Service Procedure:
1. Clear the PPI memory by pressing the PPI Memory Reset
Switch (S1) on the rear connector panel, then remove and
reapply power to the injector.
2. If the condition occurs repeatedly, replace the PPI card.