SPECIFICATIONS
20
DLRO 10 DLRO 10X
Maximum Lead Resistance 100 mΩ total for 10A operation irrespective of battery condition.
Voltmeter Input Impedance > 200 kΩ
Hum Rejection Less than 1% ± 20 digits additional error
with 100 mV peak 50/60 Hz. on the potential leads.
Warning will show if hum or noise exceeds this level.
Storage 700 tests
Memo Field Up to 200 characters per test via
integral alphanumeric keypad.
Battery Capacity 7 Ah NiMH rechargeable.
Life Typically 1000 x 10 A tests before recharge
Recharge Via External 90V - 260 V 50/60 Hz charger or from 12 to 15 V d.c. supply
Charging Time 2.5 hours to 90% capacity, 4 hrs for full charge
Temperature Operation +5°C to +45°C (41°F to 113°F) at full specification.
-10°C to +50°C (14°F to 122°F) at reduced acuracy
Storage -30°C to +70°C (-22°F to 158°F)
Calibration 20°C (68°F)
Coefficient <0.01% per oC from 5°C to 40°C (<0.006% per °F from 41°F to 104°F)
Standard charging 0°C to +45°C (32°F to 113°F)
Fast charging +10°C to +45°C (50°F to 113°F)
Humidity Max 90% RH @ 40°C (104°F) non-condensing
Altitude Max 2000m (6562 ft.) to full safety specifications
Safety In accordance with IEC61010-1 600 V Category III - only when DH6 leads are used
EMC In accordance with IEC61326-1
Operation inaccuracies Refer to www.megger.com
Dimensions 220 x 100 x 237 mm (8.6 x 4 x 9.5 in)
Weight 2.6 kg (5 3/4 lb.) including battery module