Display: 3
3
4-digits 4000 counts or 4
3
4-digits 40,000
counts selectable (5 digits 99999 counts for Hz), and 4-
digits 9999 counts dual display LCD
Polarity: Automatic
Update Rate: 3
3
4-digit Data: 5 per second nominal; 4
3
4-
digit Data: 1.25 per second nominal; 43 Segments Bar
graph: 128 per second max
Low Battery: The indicator appears when the battery
voltage drops below approx. 5.8V
Operating Temperature: 0°C to 35°C, 0-80% R.H.;
35°C to 50°C, 0-70% R.H.
Storage Temperature: -20°C to 60°C, 80% R.H.
(with battery removed)
Temperature Coefficient: Nominal 0.15 x (specified
accuracy)/°C @ (0°C-18°C or 28°C-50°C), or otherwise
specified
Power Supply: Single Alkaline 9V battery NEDA1604A,
JIS6AM6 or IEC6LF22
APO Timing: 4.5 minutes/ (default)/55/disabled
APO Consumption: 20µA typical
Sensing: True RMS conversion for M8037; Average
responding for M8035
Power Consumption: 12 mA typical
Weight: 390g; 500g with holster.
Dimensions: 186mm x 87mm x 35.5mm;
198mm x 97mm x 55mm with holster.
Safety: The instruments meet the requirements for
double insulation to IEC 1010-1 (1995), EN 61010-1
(1995) to:
Terminal V/R: Installation category III**, 1000V d.c.
and 750V a.c.
Terminal mA/µA:Installation category III**, 500V a.c.
Installation category II*, 250V d.c.
Terminal A: Installation category III**, 600V a.c.
Installation category II*, 250V d.c.
EMC: In accordance with IEC61326 including
amendment No.1