EasyManua.ls Logo

MELAG Vacuklav 40 B+ Evolution - Technical Data

MELAG Vacuklav 40 B+ Evolution
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 Technical data
81
14 Technical data
Device type Vacuklav 40 B+ Vacuklav 44 B+
Device dimensions (H x W x D) 56.5 x 46 x 58 cm 56.5 x 46 x 69 cm
Empty weight 55 kg 64 kg
Operating weight 66 kg 76 kg
Sterilization chamber
Chamber diameter/depth Ø 25 cm | 35 cm Ø 25 cm | 45 cm
Chamber volume 18.4 l 23.8 l
Electrical connection
Electricity supply 220-240 V, 50/60 Hz, 3400 W
max. voltage range 207-253 V
Building fuses 16 A, RCD 30 mA
Length of the power cable 2 m
Overvoltage category (in
accordance with EN 61010-1)
Transient overvoltage up to the values of overvoltage category II
Degree of air pollution (in
accordance with EN 61010-1)
2
Surrounding conditions
Noise emission 72 dB(A)
Heat emission (with max. load) 0.95 kWh
Ambient temperature 5-40 °C (ideal range 16-26 °C)
Degree of protection
(following IEC 60529)
IP20
Relative humidity max. 80 % at 31 °C, decreases in a linear fashion up to max. 50 %
relative humidity at 40 °C
Max. altitude 4000 m
Installation location Interior of a building
Cooling water connection
Water quality drinking water
Min. flow pressure 0.5 bar at 3 l/min
Max. static water pressure 10 bar
Max. water temperature max. 20 °C (ideal 15 °C)
Max. water consumption c. 30 l c. 35 l
Feed water connection
Water quality EN 13060, Appendix C
Min. flow pressure 1.5 bar at 3 l/min
Min. static water pressure 2 bar
Max. static water pressure 10 bar
Max. water consumption 0.75 l 0.83 l
Wastewater connection
Max. through-flow volume 3.5 l 3.5 l
Max. water temperature 80 °C 80 °C

Table of Contents