EasyManua.ls Logo

Merck Millipore ReadyStream - Priming of the filter set

Merck Millipore ReadyStream
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
OR
When the ReadyStream
®
media bag is connected to the
instrument, scan the label.
If a new bag is used, the media type (e.g. BPW100L),
the maximum number of dispensations (e.g. 29) of
a dened volume (e.g. 3375 mL), and the lifetime
after reconstitution (e.g. 5 days) are displayed. This
information is stored in the instrument.
If the media bag has been previously used on the same
instrument, the remaining volume of media (e.g. 20 x
225 mL), and the remaining lifetime (e.g. 1 day) are
displayed.
“Filter” and “Bag” LEDs should be green at this stage.
During the process, pressing [CANCEL] will
restart the bag setup.
Always use the same media bag on the same
ReadyStream
®
instrument.
If the shelf life of the media bag is exceeded or if
a wrong barcode is scanned, an error message will
be displayed, and the “Bag” LED will blink red for 3
seconds and the user will be asked to scan and connect
a media bag.
Priming of the lter set
For optimal performance, any air in the inlet tubing, in
the instrument, and in the lter set must be removed.
Ensure the media unit is connected to a water
source of sucient ow rate or, alternatively,
to a large container of deionized water.
During priming, excess water will be
dispensed through the gray dispenser on
the dispensing unit. Place a bag in the bag
holder or position a 1 L container below the
dispenser to collect the water.
Following the media bag installation on the
instrument, the user is prompted to prime the lter.
To remove air trapped in the lter, open the vent port on
the lter by turning slightly (one turn counter- clockwise).
Press the [PRIME] button, indicated by the “Prime”
LED blinking orange.
For more ecient priming, open the vent
before pressing [PRIME]. Close the vent when
the water level reaches the vent.
Excess water dripping from the vent will be
retained in the front door. Alternatively place
a clean tissue around the vent to collect the
droplets.
!
!
!