27
Storing bag with reconstituted
media for later usage
It is possible to remove a fully reconstituted media bag
with its lter set connected for later usage (e.g. when
several bags are prepared in advance, or when the bag
needs to be removed for cleaning the instrument).
Before disconnecting the media bag and lter from
the system, ush the tubing with 1 L of water (e.g.
by scanning the attached datamatrix and pressing on
the [WATER] button). Carefully recap the outlet with
a clean orange cap before disconnecting the lter set
from the media unit and close the lter set with the
blue cap.
Ensure the stopcock actuator is in home
position before removing the tubing set from
the instrument.
Recap the outlet with the orange cap before
disconnecting the lter from the instrument
to prevent backow from the lter set inlet
connector.
Store the media bag with the lter attached
at conditions mentioned in the media bag
datasheet.
After reconnection of the media bag and the lter with the
system, follow the process steps previously described to
setup the consumables, repeating at least ve times the
priming step to ush the tubing with approximatively 1 L of
water.
If the media bag and the lter set in use
are reconnected to another instrument, the
tracking data will not be accurate. Please
make sure to have enough remaining media in
the bag before starting to dispense.
Priority to ow rate/priority
to temperature
The ReadyStream
®
system can deliver culture media
or water in two modes: either with a priority to
temperature or with a priority to ow rate (speed).
Pressing the [SPEED] button switches between those
two modes.
The active priority is indicated by an LED on the front
panel, and the non-priority parameter is displayed in
brackets.
When a temperature parameter is requested
in ow-rate-priority mode, the system
will try to come as close as possible to the
set temperature. The ability to reach the
temperature is linked to your country’s power
limitation and inlet water temperature. The
±2 °C precision is not guaranteed is this mode.
!
!
!
1. Recap outlet
with orange cap
2. Recap lter inlet
with blue cap