30
Maintenance
Cleaning the surface of the instrument
The external surface of the instrument can be cleaned
using a regular cleaning and decontamination agent.
We recommend avoiding to directly spray on surfaces
and any active moving parts. It is preferable to use a
wipe that has been pre-impregnated with the cleaning
agent.
Do not use bleach or aggressive chemicals to
clean the surface.
Do not autoclave the plastic parts.
The following reagents are recommended external
cleaning:
Reagent External cleaning
Quaternary ammonium Yes
Isopropanol Yes
Ethanol Yes
Sanitization of the internal uidic path
If the system has not been used regularly for more
than 2 weeks, it is recommended to sanitize the
instrument using a specic procedure.
!
Volume
(mL)
Temperature
(°C)
Concentration
(x)
Flow
rate
(L/min)
Priority
mode To scan
90 No heating 1 1 Flow rate
100 No heating 1 1 Flow rate
225 No heating 1 1.3 Flow rate
225 No heating 1 1.5 Flow rate
225 36 1 1 Temperature
3375 No heating 1 1.7 Flow rate
3375 36 1 1 Temperature
Common test data matrices
The following data matrices can be scanned to
congure the instrument in routine dispensing.
Follow instructions on page 25 on how to modify the
parameters and on how to store them as preset on the
system. Custom data matrices can also be generated
as described in the application sheet “ReadyStream
®
creating custom data matrix codes”.