47
AQUALAB VSA
1. Before starting a test, set the Volatiles: Yes/No option in the Configuration menu. If the
sample has volatiles, select Yes otherwise select No.
2. On the Main screen, select the Test Wizard.
The Warning screen appears to let the user know that any data points currently on the
instrument from previous tests disappear when a new test begins.
Figure 37 Warning screen before new test
3. Select the double right arrow to continue.
4. A new message reminding the user to make sure the water chamber is full and there is
still active (blue) desiccant in the desiccant tubes.
Section4 has instructions on filling the water chamber and Section5.1 has instructions
for replacing the desiccant.
Figure 38 Message reminding user to check water and desiccant levels
5. Press the double right arrow to continue or the double left arrow to go back.
The Edit Test screen appears. The name of a previous test appears by default.