MI 3201 TeraOhm 5 kV Plus Warnings
10
Do not touch any conductive parts of equipment under test during the test.
Make sure that the tested object is disconnected (mains voltage disconnected)
before starting the Insulation Resistance measurement!
Do not touch the tested object whilst testing it, risk of electric shock!
In case of a capacitive test object (long tested cable etc.), automatic discharge
of the object may not be done immediately after finishing the measurement –
“Please wait, discharging” message will be displayed.
HANDLING WITH CAPACITIVE LOADS
Note that 40 nF charged to 1 kV or 9 nF charged to 5 kV are hazardous live!
Never touch the measured object during the testing until it is totally
discharged.
Maximum external voltage between any two leads is 600 V (CAT IV
environment).