MI 3201 TeraOhm 5 kV Plus Measurements
36
5.8 Withstanding voltage
This function offers Withstanding Voltage test of insulation material. It covers two types
of tests:
a) Breakdown voltage testing of high voltage device, e.g. transient suppressors
b) DC withstanding voltage test for insulation coordination purposes.
Both functions require breakdown current detection. In the function, the test voltage is
increased from the starting voltage to the stop voltage over a predefined time (set by the
parameters). The Stop voltage is then maintained for a predefined test time, (see the
Fig. 28).
Ut Ut
Ustop
Ub
Ustart Ustart
U
step
00
T
step
T
end
tt
Fig. 28. Test voltage presentation without breakdown (left part) and with breakdown
(right part)
Ut........ Test voltage
Ustop .. End test voltage
Ustep .. Voltage step approx. 25 V (fixed value - not presetable)
Ustart.. Starting voltage
Tstep... Test voltage duration per step
Tend ... Constant test voltage duration after reaching End value
t........... Time
Ub....... Breakdown voltage
Selecting this function displays the following states. Fig. 29 shows the initial screen and
a screen with results after the completion of a measurement
Initial display
Display with results