3.2. Remote control commands
794 Basic Titrino
98
&Assembly
Object Description Input range Reference
& Root
:
Assembly Assembly control
. Bur Buret
.Rates Rates
.Forward Forward rate
.Select Type of rate control digital, analog 3.2.2.83.
.Digital Digital rate 0...150, max. ditto
.Reverse as for forward rate
.Select Type of rate control digital, analog ditto
.Digital Digital rate 0...150, max. ditto
.Fill Fill $G,$H,$C 3.2.2.84.
.ModeDis Dispensing $G,$S,$H,$C 3.2.2.85.
.Select Type of dispensing control volume, time ditto
.V Volume to be dispensed 0.0001...0.1...9999 ditto
.Time Time to dispense 0.25...1...86 400 ditto
.VStop Limit volume 0.0001...9999, OFF ditto
.AutoFill Filling after each increment ON, OFF ditto
.Meas Measuring
.Status Measuring ON/OFF ON, OFF 3.2.2.86.
.MeasInput Selection of measuring input 1, 2, diff., Ipol, Upol, Temp ditto
.Ipol Polarization current 0...1...±127 ditto
.Upol Polarization voltage 0...400...±1270 ditto
.Outputs I/O outputs
.AutoEOD Automatic output of EOD ON, OFF 3.2.2.87.
.SetLines Set I/O lines $G ditto
.L0 Signal on L0 active,inactive,pulse,OFF ditto
up to L 3
.ResetLines Reset I/O lines $G ditto