EasyManua.ls Logo

Metrohm 794 Basic Titrino - Page 124

Metrohm 794 Basic Titrino
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.2. Remote control commands
794 Basic Titrino
120
Assembly.Outputs.SetLines $G
Assembly.Outputs.SetLines.L0 active, inactive, pulse, OFF
up to .L 3
Assembly.Outputs.ResetLines $G
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 162.
Warnings:
L2 is the EOD line. If you have "AutoEOD" set to "ON", an active
line 2 is set to "inactive" by the EOD pulse.
L3 is the line of the activate pulse. An active line 3 is set to "in-
active" by the activate pulse.
.ResetLines: Lines are set to the inactive status (= high).
3.2.2.88. Setup.Keycode ON, OFF
ON means the key code of a key pressed on the Titrino is outputted. The key
code comprises 2 ASCII characters; table of the keys with their code, see
page 139. A keystroke of key 11 is sent as follows:
#11
The beginning of the message is marked by a space (ASCII 32).
3.2.2.89. Setup.Tree.Short ON, OFF
Setup.Tree.ChangedOnly ON, OFF
Definition of the type of answer to $Q.
.Short: With "ON", each path is sent with only the necessary amount of
characters in order to be unequivocal (printed in bold in this manual). A com-
bination of .Short and .ChangedOnly is not possible.
.ChangedOnly: Sends only the changed values, i.e. values which have
been edited. All paths are sent absolute, i.e. from the root.
3.2.2.90. Setup.Trace ON, OFF
The Titrino automatically reports when a value has been confirmed with <en-
ter> at the Titrino. Message, e.g.:
&SmplData.OFFSilo.Id1"Trace"
The beginning of the message is marked by a space (ASCII 32).
3.2.2.91. Setup.Lock.Keyboard ON, OFF
Setup.Lock.Config ON, OFF
Setup.Lock.Parameter ON, OFF
Setup.Lock.SmplData ON, OFF

Table of Contents