4.1 Error and special messages
794 Basic Titrino
135
stop V reached The determination has been stopped as the stop volume has
been reached.
system error 3 The instrument adjustment data have been overwritten.
Exit: <CLEAR>. Default adjustment data are set. The error
message appears each time the instrument is switched on
until it has been readjusted (Metrohm service).
TIP terminated TIP has been terminated.
wrong sample With SET, with preset titration direction the first measured
value is outside the end point.
Error messages in connection with the data transfer
If neither a computer nor a printer is attached, the report output at the end of the titration
must be switched off:
Receive errors:
E36 Parity
Exit: <QUIT> and set corresponding quantity the
same on both instruments
E37 Stop bit
Exit: <QUIT> and set corresponding quantity the
same on both instruments
E38 Overrun error. At least 1 character could not be read.
Exit: <QUIT>
E39 Overflow of the receive buffer of the Titrino (> 82
characters).
Exit: <QUIT>
Send errors:
E40 DSR=OFF
E41 DCD=OFF
E42 CTS=OFF
Handshake unsatisfactory for more than 1 s.
Exit: <QUIT> Is the receiver switched on and ready
to receive?
E43 The transmission of the Titrino has been interrupted
with XOFF for at least 3 s.
Exit: <QUIT>.