2.8 Result calculations
794 Basic Titrino
53
RS1 text RS1
RS1 decimal places 2
RS1 unit: %
Text for result output (up to 8 characters)
Text input see page 8.
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/L, mg/mL, mol/L,
mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to
6 characters).
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00 Sample size, see page 69.
C01...C19 Method specific operands, see page 54. They are stored with the
method in the method memory.
C21...C23 Sample specific operands, see page 69ff.
C26, 27 Mean values from silo calculations.
C30...C39 Common variables.
C40 Initial measured value of the sample, last measured value in MEAS.
C41 End volume.
C42 Determination time.
C43 Volume drift for SET with conditioning.
C44 Temperature.
C45 Dispensed start volume.
C46 Asymmetry-pH (calibration).
C47 Electrode slope (calibration).
C51...C59 Fix EP for DET and MET.
C61...C69 pK/HNP values for DET and MET.
C70...C79 Temporary variables for calculations in TIP.