3.2 Remote control commands
794 Basic Titrino
93
&Config
Object Description Input range Reference
& Root
:
├ Config Instrument configuration
├ .PeriphUnit Selection of peripheral units
│ ├ .CharSet1 External printer Epson,Seiko,Citizen
│ │ IBM,HP 3.2.2.48.
│ ├ .Balance Selection of balance Sartorius,Mettler,Mettler AT
│ │ AND,Precisa 3.2.2.49.
│ ├ .Plot Selection of plot at analog output U, dU/dt, V, dV/dt
│ │ U(rel), T ditto
│
├ .Aux Miscellaneous
│ ├ .Language Dialog language english, deutsch,
│ │ francais, espanol, italiano,
│ │ portugese, svenska 3.2.2.50.
│ ├ .Set Setting of date and time $G 3.2.2.51.
│ │ ├ .Date Date XXXX-XX-XX
│ │ ├ .Time Time XX:XX
│ ├ .RunNo Run number 0...9999 3.2.2.52.
│ ├ .AutoStart Automatic start 1...9999, OFF 3.2.2.53.
│ ├ .StartDelay Start delay time 0...999 999 3.2.2.54.
│ ├ .DevName Device label 8 ASCII char. 3.2.2.55.
│ ├ .Prog Program version read only 3.2.2.56.
│
├ .RSSet Settings RS232 $G 3.2.2.57.
│ ├ .Baud Baud rate 300,600,1200,2400,4800,
│ │ 9600
│ ├ .DataBit Number of data bits 7, 8 ditto
│ ├ .StopBit Number of stop bits 1, 2 ditto
│ ├ .Parity Parity even, odd, none ditto
│ ├ .Handsh Handshake HWs, HWf, SWchar,
│ │ SWline, none ditto
│
├ .ComVar Values of common variables
│ ├ .C30 C30 0... ±999 999 3.2.2.58.
│ ├ up to C39 0... ±999 999
│