2.3 Composition of a method
Metrohm 838 Advanced Sample Processor, Operation 27
2.3.2 Sample Processor settings
parameters
>changer settings
Submenu for setting the Sample Processor functions
Open the submenu with <ENTER>
>changer settings
rack name: *
The sample rack assigned to the method
*,10 ASCII characters
* = no particular rack
With this setting the user can be forced to use a particular rack for the se-
lected method. If this is not required then
rack name: * must be selected.
>changer settings
lift rate T1 25 mm/s
Vertical movement rate of the lift
5…25 mm/s
>changer settings
shift rate 20°/s
Speed at which the rack rotates
in degrees per second
3…20
>changer settings
shift direction: auto.
Direction of rotation of the sample rack
+,–,auto. auto. = the Sample Processor selects the shortest
path for the rotation itself.
>changer settings
rotat. increment: 5.00°
Increment angle for the command
MOVE +/-rotate
0.00…5.00…270.00°
>changer settings
swing rate T1 55°/s
Swing rate of the robotic arm in de-
grees/second
10…55
>changer settings
swing increment 10.00°
Increment angle for the command
MOVE +/-swing
0.00…10.00…180
>changer settings
on beaker error: MOVE
Definition of reaction if beaker is missing
Not implemented.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 838 Advanced Sample Processor and is the answer not in the manual?

Metrohm 838 Advanced Sample Processor Specifications

General IconGeneral
BrandMetrohm
Model838 Advanced Sample Processor
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals