2.8 Dosing and liquid handling
Metrohm 838 Advanced Sample Processor, Operation 61
2.8.5 Liquid handling functions in detail
Dosing
11
Example:
DOS: 1.1 : 1.000 mL Default port = 1
'Normal' dosing can be carried out by entering a volume. Automatic
cylinder filling is triggered neither before nor after the dosing process.
If the dosing piston reaches 'max. Volume' mark (10'000 impulses) dur-
ing the dosing process then the cylinder will be refilled.
Dosing a negative volume
22
Example:
DOS: 1.2 : -1.000 mL Default port = 1
If a negative value is set for a volume then the dosing process will take
part in the opposite direction, i.e. a liquid will be aspirated through the
given port. Automatic cylinder filling is triggered neither before nor after
the dosing process.
If the dosing piston reaches the zero mark during the dosing process
then the cylinder will be refilled.
Do not select any volume that is larger than the nominal cylinder vol-
ume. Aspiration should take place with a single piston stroke.
This function can be used for pipetting.
Filling
33
Example:
DOS: 1.3 : fill mL Default port = 2
Filling the cylinder can be carried out from a freely selectable port. After
the filling process the stopcock remains at the selected port.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 838 Advanced Sample Processor and is the answer not in the manual?

Metrohm 838 Advanced Sample Processor Specifications

General IconGeneral
BrandMetrohm
Model838 Advanced Sample Processor
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals