3.6 Error messages
Metrohm 838 Advanced Sample Processor, Appendix 77
SCAN timeout The connected device has not transmitted the expected sig-
nal within the defined timeout time. The sample determina-
tion may not have been carried out properly, or the connec-
tion has been interrupted. Check the connected device.
* service recommended The warning limit of the running time meter has been
reached. It is time for the Sample Processor to be serviced.
Please contact your local Metrohm service department.
user memory full The memory for the user-defined methods is full. Before new
methods can be stored methods that are not used, or only
used infrequently, must be deleted.
wrong rack The rack placed on the turntable is not the rack assigned to
the method under
parameters.
trap error xxx Unexpected program error; switch instrument off and then on
again.
No display, LEDs Tower 1
and Tower 2 light up LCD error (system error 7). Contact Service.