4 Operation
892 Professional Rancimat
■■■■■■■■
33
NOTICE
In order to carry out at least one double determination on the 892
Professional Rancimat, you have to extract approx. 10 g pure oil or
fat, allowing for a certain transfer loss (for more than two determi-
nations, an accordingly higher amount is necessary).
Prepare a clean 250 mL round-bottom flask with standard
ground-joint and a folded filter.
Filtrate, if possible protected from light. Rewash the residue with a
little petroleum ether.
NOTICE
If the folded filter becomes blocked (e.g. due to the consistency of
the sample material), a Soxhlet apparatus should be used for solid/
liquid separation. You can use up to 40 g of sample material per
batch.
3
Distillation and filtration
Distill the petroleum ether from the clear, perhaps slightly yellow
extract.
NOTICE
It is safest and easiest if you do this in a rotary evaporator; in a
slight vacuum, protected from direct light (cover the water bath
e.g. with aluminum foil) and at a temperature of +30 - 35 °C the
petroleum ether can gently and efficiently be removed.
Dry the oil/fat sample after the distillation has been completed for
approx. 30 min at a pressure of < 1,330 Pa (13.3 mbar).
Filtrate the oil/fat sample now present together with water-free
Na
2
SO
4
with a folded filter. If necessary, work in the drying oven
at a temperature exceeding the melting point of this fat by 10 °C.
4
Further sample preparation
Treat the isolated oil/fat samples afterwards just as pure oils and
fats.