4.3 Determinations
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
892 Professional Rancimat
4
Connecting and inserting reaction vessels
NOTICE
The temperature defined in the method has to be reached before
you insert the reaction vessels, i.e., the temperature display has to
be green.
Seal the unused positions with stoppers or empty reaction vessels
for protection against impurities.
Connect the white silicone tubing which is connected to the reac-
tion vessel covers to the M8/olive tubing adapter of the measur-
ing vessel covers.
Screw the pieces of FEP tubing 250 mm to the M8/M6 thread
adapters of the reaction vessel covers and the air supply connec-
tors of the 892 Professional Rancimat.
Insert the prepared reaction vessels into the recesses of the heat-
ing block.
Press the start button to start the data recording immediately after
inserting each single reaction vessel.
4.3.4 Cleaning the accessories
1
Cleaning the measuring vessels and accessories
Clean used measuring vessels after pouring off the measuring
solution with ethanol or 2-propanol (do not use acetone!) or in
the dishwasher.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Thoroughly rinse with distilled water.
Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and
thoroughly rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Remove the protective ring to facilitate cleaning of the electrodes.
2
Cleaning the reaction vessels and accessories
Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.