7.5 Gas flow regulation
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
892 Professional Rancimat
7.5 Gas flow regulation
Pump Membrane pump (brushless motor)
Volumetric flow
range
1 - 25 L/h at 25 °C and 1,013 mbar
Maximum devia-
tion from the set
range
±(0.25 L/h + 5% of the measured value)
7.6 USB interface
USB connector
USB plug type B
7.7 Power socket
Nominal voltage
range
100 - 120 V and 220 - 240 V ± 10% (autosensing)
Frequency 50 and 60 Hz (autosensing)
Power consump-
tion
450 VA
max
Fuse Diameter 5 mm, length 20 mm
4.0 ATH (slow-acting)
7.8 Ambient temperature
Nominal function
range
+5 - +45 °C
(at 20 - 80% humidity)
Storage –20 - +70 °C
Transport –40 - +70 °C