EasyManua.ls Logo

Metrohm 913 - Detailed Menu Dialogs

Metrohm 913
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
912/913/914 Meter
4.7 Menu dialogs
4.7.1 Parameters pH/U/T and Parameters pH/U/T IS
The Parameters pH/U/T menu dialog for the parameters Measurement
and Calibration is shown below with the structure and the description.
4.7.1.1 Measuring parameters
Measuring parame-
ters
Menu dialog for the Measuring parameters.
Measuring mode
Selection dialog for selecting the measuring mode.
pH
The pH value is output.
U
The potential value is output in mV.
T
The temperature is output in °C.
Sensor name
Selection dialog for selecting a sensor from the sensor list.
In instruments with iTrodes, this is only a display field.
Order number
Display field with the sensor's Order number.
Serial number
Display field with the sensor's Serial number.
Temperature
Editing dialog for the manual entry of the measuring temperature.
Default value: 25.0 °C / input range: –999.9 - +999.9 °C
Does not apply for instruments with iTrodes.
Delta measure-
ment mV
Selection dialog
on: with the input field for the Reference
with default value: 0.0 mV / input range: –1500.0 - +1500.0 mV
off: is the default value
4.7.1.2 Calibration parameters
Calibration param.
Menu dialog for the Calibration param..
Temperature
Editing dialog for manually entering the calibration temperature.
Default value: 25.0 °C / input range: 0.0 - 99.9 °C
Report
Selection dialog
on
off: is the default value

Table of Contents

Related product manuals