4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
912/913/914 Meter
pH
Slope upper
limit
Editing dialog for entering the upper limit value.
Default value: 103.0 / input range: 1.0 - 999.9%
pH
Limit value
pH(0)
Selection dialog for selecting whether the limit value is to be applied.
on
off: is the default value
pH
pH(0) lower limit Editing dialog for entering the lower limit value.
Default value: 6.40 / input range: 0.00 - 99.99
pH
pH(0) upper
limit
Editing dialog for entering the upper limit value.
Default value: 8.00 / input range: 0.00 - 99.99
K
Limit value c Selection dialog for selecting whether the limit value is to be applied.
on
off: is the default value
K
c lower limit Editing dialog for entering the lower limit value.
Default value: 0.400 /cm / input range: 0.001 - 500 /cm
K
c upper limit Editing dialog for entering the upper limit value.
Default value: 0.550 /cm / input range: 0.001 - 500 /cm
DO
Temperature off-
set
Editing dialog for entering/modifying the Temperature offset.
Default value: 0.0 °C
Input range: –5.0 - +5.0 °C
DO
LED intensity
(%)
Selection dialog for selecting the LED intensity (%).
10 %
20 %: is the default value
30 %
DO
Cap type Display field for the Cap type.
e.g. MA7-530-200
DO
Limit value dphi
100 %
Selection dialog for selecting whether the limit value is to be applied.
on: is the default value
off
DO
lower limit Editing dialog for the lower limit.
Default value: 15 ° / input range: 15.0 - 30.0 °