EasyManua.ls Logo

METRON QA-1290 - Specifications

METRON QA-1290
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test results, shown in the QA-1290's digital display, can be printed
out directly, or transferred to a PC via the ansur QA-1290 test auto-
mation software. ansur lets you design test protocols, remotely con-
trol the QA-1290, and store the test results.
1.2
Specifications
1. Simulations
Simulation technique: Oscillometric
Calibration Tables used: DINAMAP, HP-Merlin
Simulation repeatability: ±0.5 mmHg
Heart Rate: 10-200 bpm, 1 bpm steps
Heart Rate Accuracy: ±1 bpm @pp<
2 mmHg
Pulse Pressure: 0.05-5.00 mmHg
Systolic/Diastolic Pres-
sure:
As shown below
Preset Pressures
(Systolic/Diastolic (MAP), in mmHg)
Adult Neonate
60/ 30 (40) 60/ 30 (40)
80/48 (58) 80/48 (58)
100/65 (77) 100/65 (77)
120/80 (95) 120/80 (95)
150/95 (114) 150/95 (114)
200/140 (167) 200/140 (167)
255/195 (215) 255/190 (220)
Independent Systolic/Diastolic Settings
(in mmHg)
Systolic 20 to 255
Diastolic 5 to 220
Pulse Pressure 0.05 to 5.00
Preset Artifacts and Arrhythmias
Artifacts include: Bradycardia, Tachycardia,
Geriatric, Obese, Strenuous Exercise, Mild
Exercise, and Weak Pulse
Arrhythmias include: Premature Atrial Con-
traction, Premature Ventricular Contraction,
Atrial Fibrillation, and 2
nd
Degree Heart Block
Custom Patient Pattern Mode
1-2