An arbitrary patient pattern can be
downloaded via RS-232 and ansur software.
Motion/Tremor Artifact:
Available at all BP Start screens
2. Pressure / Leakage Measurements
Pressure Range\Accuracy: 0 to 500 mmHg automatic or manual inflation
Accuracy +/- 1mmHg
Units of measure mmHg, cmH20, inH2O, PSI,
kPa, mBar.
Digital Readouts: Manometer, Start Pressure, Peak Pressure,
Pressure Drop, Deflation Time and Leak Rate.
Leak Rate Range: 1 to 500 mmHg/minute
Total Leak Measurement
Time:
60 seconds (and each subsequent 60
seconds)
1.3 General Information
Display/Control:
Type: LCD
Alphanumeric format: 4 lines, 40 characters
Display control: 7 F-keys and keypad
Data Input/Outputs (2):
Parallel printer-port (1): Bi-directional 25-pin, Type
Centronics, RS-232C: (1) for Remote Control
Power:
From 110 VAC to 240 VAC, 47/63 Hz
Housing:
Aluminum
Weight (w/o batt.):
5.5 kg / 12.1 lb.
Dimensions:
Height: 9.5 cm / 4 in.
Width: 30.0 cm / 11.7 in.
Length: 27.0 cm / 11 in.
Temperature
Requirements:
+15 °C - 35 °C / +59 °F - +95 °F while operating
0°C - +50°C / +32°F - 122 °F for storage
Part No:
QA-1290 NIBP Analyzer
(P.N. 17500)
Standard Accessories:
QA-1290 User & Service Manual (P.N. 17525
BP Cuff Mandrels:
Medium Adult (95 mm 0D, 178 mm width)
and Neonate (27 mm 0D, 43 mm width)
QA-1290 Tubing Kit (P.N. 17515)
Cuff Adaptors:
1
DINAMAP, Luer, Luer-Lock, HP-Merlin,
Propaq/Marquette, IVAC (1/4" hose/barb),
Baum (5/32" hose/barb), 1/8" hose/barb
Optional Accessories:
QA-1290 Carrying Case (P.N. 17510)
Ansur plug in for QA-1290 (P.N.
1
Specifications may be subject to change without notice.
1
Note: Adapters connect to the QA-1290 Pressure Ports via quick-
disconnect, airtight, O-ring pressure fittings.
1-3