EasyManua.ls Logo

METRON QA-1290 - Configuring QA-1290

METRON QA-1290
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Main Menu Screen 1 (above) contains the following settings:
Press Equipment Code (F7) to enter the facility code for
equipment being tested. Then press Enter to save.
Press Leak Test (F6) to advance to the Leak Test screen.
Press BP Test (F5) to advance to the BP Test screens.
Press MORE (F1) to advance to Main Menu Screen 2. (See be-
low).
Over-Pressure Cut-Off Test ............> F7
Manometer ............................> F6
Print header/footer ...................> F5
GO BACK SET UP
F1 F2 F3 F4
Press Over-Pressure Cut-Off Test/Manometer (F7) to advance
to that test’s screen.
Press Manometer (F6) to advance to that test’s screen.
Press Print Header / footer (F6) to advance to the Print Screen
(see paragraph 3.6 below).
Press GO BACK (F2) to return to Main Menu Screen 1, SETUP
(F3) to advance to the Set Up Screen 1.
3.5 Configuring QA-1290
Pressing SETUP (F3) in either Main Menu Screen advances you to a
series of screens that allow you to configure QA-1290 settings and
store them for use in all testing. There are three Setup Screens.
SYSTEM SETUP F7
Operator :.........................> F6
Establishment:.........................> F5
MORE STORE MAIN MENU
F1 F2 F3 F4
Setup Screen 1. This screen (above) contains the following settings:
Press Operator (F6). Use the alphanumeric keys to enter the
test operator’s name or other identifying data. Press Enter to
save.
3-4

Related product manuals