EasyManua.ls Logo

MFJ 223 - Single-Frequency Setup; DDS Signal Generator

MFJ 223
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MFJ-223 Vector Impedance Antenna Analyzer
MFJ Enterprises, Inc
10 Version 1
circuits or matching networks. Pressing Run/Stop again toggles
the generator off.
4.2 S-Antenna Bar: The S-Antenna bar warns when your
antenna is being energized by powerful external signals from a
nearby transmitter. It also indicates when the analyzer's DDS
generator is running, but its primary function is to detect
interference when the generator is stopped. A full-scale reading
means power in excess of 0-dBm is being fed back into the
analyzer from an external source. In this event, disconnect the
analyzer immediately to avoid damaging the coupler. Although
marked in S-units, the meter is not logarithmic, so any reading
exceeding S2 or S3 could potentially interfere and cause
inaccurate readings.
Overload Warning: When testing, check the S-Antenna bar
with the generator turned off to see if disruptive RF levels are
being picked up by the antenna.
4.3 Single-Frequency Setup:
To enter the Single-
Frequency mode from the Boot Screen:
Press the Single command key. The bar-graph
(labeled Single) will appear.
Connect the test antenna to the analyzer and check
the S-Antenna scale.
Use the Encoder to select a step and tune to your first
test frequency.
Press the Run/Stop key to start the DDS stimulus
generator running.
Tune as needed to explore the antenna's
measurements of interest.
[ ] Press the Run/Stop key again to terminate the test
run.
The Run/Stop function conserves battery power, allowing the
analyzer to operate for extended periods without recharging.
4.4 DDS Signal Generator:
The analyzer's test generator
may also be used as a highly accurate RF signal generator for
aligning receivers, tuning networks, or driving amplifier chains.
Typical output power level is +5 dBm 1 dB) with 20-dB or
better harmonic suppression. Frequency resolution is 100 Hz
from 0.5 to 60.0 MHz with <5 ppm error. Anytime the analyzer
is in Single-Frequency mode and the Run/Stop key is

Related product manuals