CONTROLS & INDICATORS
The following illustrations show the location of the sterilizer’s controls and indi-
cators and the chart on the following pages describes their function.
Control Function
TEMPERATURE
REGULATOR
Controls operating temperature in chamber. A dial stop
prevents setting temperature higher than 270°F (132°C).
CYCLE TIMER Controls time that load will be exposed to steam at set
temperature. Timer will buzz at termination of cycle.
Timer is graduated in minutes from 0 to 30 and functions
as ON/OFF switch.
PILOT LIGHT (green) Illuminates when heater is energized.
TEMPERATURE GUAGE Indicates temperature in chamber. Dial is graduated from
170 to 270° F ( 77 to 132° C ).
RESET BUTTON Restores power after Low Water Thermostat has opened.
After unit has cooled, depressing the Reset Button closes
thermostat.
Controls &
Indicators
Return To Table Of Contents