Safety Features
The M7 SpeedClave™ sterilizers . . .
• have their chamber temperature monitored during a cycle to prevent an
overheat condition. If chamber temperature reaches or exceeds 146°C
(295°F), power to the sterilizer is cut off.
• have a pressure relief valve which opens at 31 psi (214 kPa) to provide
backup protection to reduce chamber pressure in event that chamber steam
pressure exceeds allowable limit. Released steam is directed out to reser-
voir.
Description