20
Venting Sterilizer
1. When cycle is complete buzzer will
sound for 1 minute. Turn timer off and
place door handle to Vent position.
2. Hold Fill / Vent lever down until door
“pops” inward and then release lever.
Leave door handle in Vent position. Do
not open door.
Drying Cycle
1. Allow door handle to remain in Vent
position for 15 minutes. Do not
open
door or turn on heat during this time
period.
2. After Drying time has elapsed, open
door and remove contents. Trays may
be placed on racks located on top of
sterilizer.
NOTE
Allowing Sterilizer to set without venting
will cause items to come out wet.
NOTE
Sterilizer makes loud gurgling sound while it vents. Opening door after vent-
ing will prevent items from drying properly. Leave door handle in Vent posi-
tion.
EQUIPMENT ALERT
Do not turn on heat or open door during 15 minute drying period.