23
Monthly
1. Flush the System - To protect intricate parts of unit, system must be flushed
once a month with Speed-Clean Sterilizer Cleaner:
(a) Drain reservoir and fill with clean distilled or demineralized water.
(b) Add one (1) ounce of Speed-Clean Sterilizer Cleaner to a cool chamber
then, fill chamber to its normal level.
(c) Run one 6 minute cycle at 270°F (132°C). Instruments must not be ster-
ilized while cleaning sterilizer.
(d) Drain cleaning solution from chamber and reservoir. Fill reservoir with
clean distilled or demineralized water and run one 3 minute cycle at
270°F (132°C).
(e) Drain reservoir and allow sterilizer to cool to room temperature and
remove trays from wire rack.
(f) Lift up on left edge of tray plate (1) until tray plate “pops” free of wire
rack (2).
(g) Hold tray plate (1) in a vertical position, and remove from chamber.
(h) Gently squeeze bottom of wire rack (2) together and remove from
chamber.
(i) Wipe out inside of chamber being careful not to damage heater element.
(j) Wipe off trays, wire rack, and tray plate..
(k) Gently squeeze bottom of wire rack (2) together and insert into chamber,
pushing rack in as far as it will go.
(l) Hold the tray plate (1) in a vertical position with the right side of the tray
plate down; then insert it into the chamber.
NOTE
Be sure to install the wire rack so that the straight ends (A) of the rack are on
the right side of the chamber, and the offset ends (B)of the rack are on the left
side of the chamber.