6
plug on sterilizer power cord must be plugged into a matching three-pronged,
grounded, isolated, correctly polarized receptacle. Check serial number label
on back panel of sterilizer to verify voltage rating for unit.
DESCRIPTION
Operational Features
M7 SpeedClave Sterilizers . . .
have a three prong plug-in operation for standard electrical outlets.
have a self-contained reservoir to hold distilled or demineralized water. No
exterior plumbing required.
have a lever for Fill / Vent located on front of unit for easy access.
will emit an audible signal when time on cycle timer has elapsed.
after manually venting steam pressure from chamber, door handle is placed
in middle vent position to facilitate drying.
have a Cycle Timer to control amount of time that load is exposed to steam
at set temperature.
have a Temperature Regulator that turns heating element On and Off to
control operating temperature in chamber.
have a Pilot Light that illuminates whenever heater is energized.
have a Low Water Cut-Out Switch to prevent operation when water level is
low.
have a Reset Button to restore operation after Low Water Cut-Out Switch
has tripped and water has been added.
have a drain coupler located on front for easy draining of self contained res-
ervoir.
Mechanical Features
The M7 SpeedClave sterilizers . . .
have a tray rack which can hold one small and two large trays. The tray rack
can be removed for periodic cleaning of the chamber, tray plate, and tray
rack.
have a Temperature Gauge to indicate internal chamber temperature. Dial
is graduated from 170 to 270°F (77 to 132°C).
have a drain coupling, adapter and hose assembly allows draining reservoir
from front of unit.
Description
Installation
Return To Table Of Contents