Set Time and Temperature.
1. Set Timer (1) to approximately 15
minutes. Pilot Light (2) illuminates to
show heater is on.
2. Turn Temperature Regulator (3)
knob fully to left (counterclock-
wise). This is maximum setting of
270° F (132° C).
3. When Temperature Gauge (4) reaches desired temperature then, reset
Timer (1) for desired sterilization time. Pilot Light (2) will flash on and off dur-
ing sterilization cycle.
NOTE
To set proper time and tempera-
ture refer to chart on top of ster-
ilizer or to Recommended
Temp. & Times (Page 15).
NOTE
If desired temperature is 250° F (121° C) instead of 270° F (132° C) when
Temperature Gauge (4) reaches 250° F (121° C) immediately
turn Tempera-
ture Regulator knob (3) slowly
clockwise until pilot lights (2) goes out.
Place a piece of tape or a mark on front face of sterilizer to mark position of
knob for future reference.