INSTALLATION
Location Requirements For Sterilizer
Adherence to the following recommendations for location of the sterilizer will
contribute to optimum performance of the unit:
Support Surface (A)
- Sterilizer must be placed on a level surface to ensure that
the chamber will fill correctly. Improper water level in the chamber could cause
a sterilizer malfunction.
Support Surface Material (B)
- Formica, stainless steel, or other water and heat
resistant material. Heat from bottom of sterilizer could reach 71°C (160° F) on
the support surfaces for short periods of time.
DANGER
Do not operate this sterilizer in areas where flammable anes-
thetics are used or stored. An explosion could occur, caus-
ing personal injury.