Support Surface Depth (C) - Support surface should be approximately 41.9 cm
(16.5 in) deep. Allow at least 5 cm (2 in.) clearance behind sterilizer for air cir-
culation.
Distance To Side Wall (D)
- If sterilizer is located next to side walls, there should
be no less than 6.4 cm (2 1/2 in) clearance between side of sterilizer and wall.
Distance Above Sterilizer (E)
- If sterilizer is to be located beneath wall cabinets
or shelves, underside of cabinets or shelves should be at least 63.5 cm (25 in)
above support surface, to provide access to top inspection cover and to allow
for changing printer paper roll.
Overhang (F)
- If sterilizer is to be located beneath an overhang, underside of
overhang should project no further than 38 cm (15 in) over rear of the sterilizer.
Location On Support Surface
- Front of sterilizer should be located near front of
support surface so water can be easily drained from tube into a container.
Neighboring Materials and Equipment
- If sterilizer will be operated in continu-
ous cycles, locate sterilizer where excessive emission of steam will not damage
materials or equipment in surrounding area.
Re-Location Requirements For Sterilizer
1. Disconnect power cord from electrical outlet and allow sterilizer to cool.
2. Drain water from reservoir or take care not to tip sterilizer, which will allow
water to spill from reservoir.
Electrical Requirements
Electrical rating for 230 VAC unit is 230 VAC, 50/60 Hz, 5 amps. Electrical rat-
ing for 120 VAC unit is 120 VAC, 60 Hz, 10 amps. Three-pronged grounding
WARNING
Use 220 - 240 VAC, 50/60 HZ alternating current only for 230
VAC models and 90 - 110 VAC, 50/60 HZ alternating current
only for 100 VAC models. Failure to do so could result in electrical
shock to personnel and will result in damage to sterilizer.
Do not use this sterilizer in an explosive or oxygen-enriched atmo-
sphere. Failure to do so could result in serious personal injury or
death.
NOTE
Grounding reliability can only be achieved if this unit is connected to a match-
ing three-pronged, grounded, isolated, correctly polarized receptacle.