22
Weekly
1. Clean Chamber and Trays - Steril-
izing chamber and instrument trays
should be cleaned weekly.
Drain water from reservoir
- A drain
coupling (1) is located on front of
unit. Insert adapter / hose assem-
bly (2) into coupling and drain into
container or sink.
Remove adapter / hose assbly. (2)
Depress release lever (3) to disen-
gage adapter while pulling outward.
Wash inside of chamber and trays
Use mild soap or Speed-Clean and
distilled or demineralized water.
Refill reservoir
Use distilled or demineralized
water only.
EQUIPMENT ALERT
Do not use abrasive or bleaching agents in chamber (i.e. steel
wool, scouring powder, bleach, etc.). Also, never
use a wire
brush. If these materials are used, possible damage to metal surfaces of
chamber and other components could result.
NOTE
If drain coupling (1) drips after in-
serting adapter / hose assembly
(2) remove and reinsert adapter /
hose assembly several times.
This will clean seals, stopping the
leak.
Operator
Maintenance
Return To Table Of Contents