OPERATOR MAINTENANCE
It is responsibility of user to establish a periodic maintenance procedure to as-
sure correct operation of equipment and reliable sterilization of loads. Contact
your local distributor or representative to develop a program for planned mainte-
nance.
Daily
1. Clean External Surfaces - Wipe with a soft dry cloth and wash occasionally
with a damp cloth and mild soap or detergent.
2. Clean Sterilizer Door Gasket
-
Clean door gasket (1) and mating
surface (2) with a damp cloth.
Examine gasket (1) for possible
damage that could prevent a
good sealing surface.
WARNING
Make sure that unit is cool when cleaning door gasket and
any mating surfaces. Failure to do so could result in serious
burns to hands.