14
Pouch or wrap items to preserve sterility after processing. Use only protec-
tive coverings designed and recommended for use with steam sterilization.
Do not wrap items too tightly. Steam penetration will be affected if an item
has excessive wrapping.
Do not stack trays on one another. Using Midmarks standard tray rack and
trays provides proper tray spacing.
Position loads on trays with appropriate space between items. Adequate
space is required between items to allow proper steam flow and drying.
Place unwrapped items on a towel.
Listed below are maximum recommended loads for each tray:
Packs to have a minimum of 1/4 in. (6.3 mm) space between each other and
away from all sterilizer surfaces.
Maximum Capacities
Load Type M7 Large Tray M7 Small Tray Sterilizer Total
Solid Items 21 instruments - 1100
grams (2.4 lbs.) or
14 instruments - 700
grams (1.6 lbs.) or
56 instruments - 3.6
kg (8.0 lbs) or
Packs (
‡)
1080 cu. cm up to 2.5
cm thick (66 cu. in. up
to 1 in. thick)
780 cu. cm up to 2.5
cm thick (48 cu. in. up
to 1 in. thick)
2940 cu. cm up to 2.5
cm thick (180 cu. in.
up to 1 in. thick)
Operation
Return To Table Of Contents