Ophthalmology
Ophthalmological surgical instrument
sets can be cleaned and disinfected in
a special mobile injector unit.
For rinsing, use demineralized water
with a conductivity of ~15 ZS/cm
(microsiemens per centimetre).
In addition, the rinse water must have a
low endotoxin and pyrogen content!
The upper level is equipped with
separate connections for hollow
instruments, e.g. rinse and suction
handpieces and cannulae.
Silicone holders and stoppers fitted on
the mesh tray fix and secure the
instruments to the inlet hose of the
mobile unit.
The lower level of the mobile unit is
designed to take E 441/1 inserts or
E 142 mesh trays to reprocess
instruments without a hollow body.
,
In washer-disinfectors used to
reprocess narrow lumen
ophthalmological instruments,
plastic mesh must not be used.
The mobile units for
ophthalmological surgical instrument
sets come with their own instructions
for use.