27
Mahr GmbH, Millimar C1208/C1216/C1240
5.1 Calibrating a measuring channel
i
Three standards are required:
one for the zero point, one for
the positive indication value
and one for the negative indi-
cation value. A nonimal value
adjuster (e.g. Millimar 1283
WN) may be used instead of
the standards.
1. Press the MENU key. The catalog item
"FEATURE (1)" is displayed.
2. Press the DATA key. The catalog item
"SETTING (4)" is displayed.
3. Press the MENU key again. The catalog item
"DISPLAY (4.1)" is displayed.
4. Press the DATA key again. The entry
"CALIBR (4.9)" is displayed.
5. Press the MENU key. The entry "PASSWRD"
is displayed along with seven zeroes below
it. The first zero is flashing.
6. Enter the password (1 000 000) and confirm
with START. "CHAN 1 (4.9.1)" is displayed.
7. Press the MENU key. "ZERO LO" is dis-
played.
8. Adjust the value 0 on the nominal value ad-
juster and confirm with START. „OFFSET
will be displayed for about 2 seconds and
then menu item „NEG VAL (4.9.1.1.1) togeth-
er with a numerical value.
9. Use the MENU, DATA and MASTER keys
to set the numerical value displayed to the
negative value of the nominal value adjuster
and accept it with START.
10. Set the nominal value adjuster to the nega-
tive calibration value and confirm with
START. „POS VAL (4.9.1.1.1.1) is displayed.
11. Adjust the positive calibration value as de-
tailed for the negative one (see points 9 and
10 of this description). „CHAN 1 (4.9.1) is dis-
played then.
12. A correction factor is calculated from the
measuring values and the nominal values
entered. In subsequent measurements, all of
the measuring values of the probe at C1 will
be multiplied by this factor.
i
Correction factors may lie in a
range between 0.3 and 3.0. If
the calculated factor exceeds
this range in either direction, the
previously set factor remains in
effect and the error message
"OUT LIM" is displayed.
13. If no other settings are to be made in the
catalog of functions and parameters, press
START. The current measuring value or re-
sult is displayed again.
To make further settings, use the ESC,
MASTER, DATA and MENU keys to navi-
gate to the corresponding item in the cata-
log of functions and parameters and make
the required settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millimar C 1240 and is the answer not in the manual?

Millimar C 1240 Specifications

General IconGeneral
BrandMillimar
ModelC 1240
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals