60
Mahr GmbH, Millimar C1208/C1216/1240
Setting the interface protocol on Millimar:
1. Press the MENU key. The catalog of func-
tions and parameters item “FEATURE (1)” is
displayed.
2. Use the DATA and MASTER keys to select
the setting “SETTING (4)”.
3. Press the MENU key. “DISPLAY (4.1)” is dis-
played.
4. Use the DATA and MASTER keys to select
the setting “COM (4.6)”.
5. Press the MENU key. “PROTOCL (4.6.1)” is
displayed.
6. Press the MENU key again. The interface
protocol that was last selected flashes.
7. Use the DATA and MASTER keys to select
the desired interface protocol (see above).
The following interface options are available:
OFF, ASCII, M1240, OPTORSS, OPTORSD,
MARTALK.
8. Press START once the desired interface pro-
tocol is displayed. The setting is adopted and
PROTOCL (4.6.1)” is displayed again.
9. If no further settings are to be made in the
catalog of functions and parameters, press
the START key. The standard display ele-
ments appear.
If further settings are necessary, navigate
to the appropriate point of the catalog of
functions and parameters using the ESC,
MASTER, DATA and MENU keys and carry
out the settings as required.
14.2 Selecting the data transfer format
The data transfer format determines the num-
ber of data bits and stop bits with which values
should be transferred and whether or not a par-
ity check should be performed.
1. Press the MENU key. The catalog of func-
tions and parameters item “FEATURE (1)” is
displayed.
2. Use the DATA and MASTER keys to select
the setting “SETTING (4)”.
3. Press the MENU key. “DISPLAY (4.1)” is dis-
played.
4. Use the DATA and MASTER keys to select
the setting “COM (4.6)”.
5. Press the MENU key. “PROTOCL (4.6.1)” is
displayed.
6. Use the MASTER key to select “FORMAT
(4.6.2)”.
7. Press the MENU key. The data transfer for-
mat that was last selected flashes.
8. Use the DATA and MASTER keys to select
the desired data transfer format. The follow-
ing data transfer formats are available:
8-N-1 8 data bits and one stop bit are
transferred. No parity check takes
place.
7-O-2 7 data bits and two stop bits are
transferred. A parity check takes
place and must produce an odd
value.
7-E-2 7 data bits and two stop bits are
transferred. A parity check takes
place and must produce an even
value.